Fully integrated 16-channel current amplifier for multichannel low-noise current measurements. Suitable for arrayed lipid membranes studies.
e16




Design your setup with our PCB

A PCB is provided to connect the amplifier to your experimental setup (e.g. flow cell,
sensor etc.). In addition, a custom designed interposer can be provided to interface with your setup. This flexibility ensures a highly personalized experience, adapting to diverse experimental requirements, while keeping high quality multichannel signal acquisition with very low noise.
Amplifier Architecture

Fully-parallel structure for current sensing, where each input is equipped with its own low-noise amplifier, digitizer, and voltage stimulus circuitry within a ±500 mV range. An additional DAC supplies a voltage reference stimulus with a maximum range of ±1.25 V. All 16 channels are integrated into an Elements dual-channel ASICs, while data processing is efficiently managed by an internal FPGA.
- Open input (RMS) noise:
- 0.41 pA rms @ 10 kHz
- 3.8 pA rms @ 100 kHz
- Current ranges: ±200pA (Gain 2.25GΩ), ±2nA (Gain 225MΩ), ±20nA (Gain 22.5MΩ), ±200nA (Gain 2.25MΩ)
- Parametric voltage protocols for pulse generator in the range of ± 500 mV
- Max available bandwidth: 100 kHz
- Max sampling rate: 200 ksps
- Each channel has individually configurable holding potential
- Controllable reference voltage in the range of ± 1250 mV
- Auto electrodes voltage offset fine compensation
- Digital Trigger Output
- USB 2.0 Data Interface
- 5 Vdc Power Supply
- Size & Weight: 23 x 79 x 108 mm, 250 g